Home » Baxter’s astute incontinence pad strategy has an RF problem
October 24, 2022 By Chris Newmarker
Baxter (NYSE: BAX) precocious announced an Urgent Medical Device Correction for its WatchCare incontinence absorption system.
The Oct. 21 announcement said that the WatchCare strategy was meant to comply with the astir caller vigor frequence standards. However, it appears to radiate RF that whitethorn impact different devices connected caregivers and patients.
In each but 2 of the reports of problems, interference took spot astatine a region of little than 1 meter. Examples of perchance superior problems include:
- Insulin pumps overdosing insulin owed to incorrect precocious glucose readings;
- “Phantom” incorrect fetal bosom code readings connected fetal monitors;
- Problems with telemetry devices, bladder scanners and infusion injection pumps.
Baxter has nary reports of injuries oregon decease related to WatchCare’s RF.
The institution is telling users and facilities to instantly find each affected devices and halt the usage of each WatchCare strategy accessories wherever imaginable until the disabling of RF-related functionality. Baxter continues to enactment to find the problem’s cause. Until it resolves the issue, it says to usage standard, non-RF-based incontinence absorption pads.
The Urgent Medical Device Correction covers Centrella Bed with WatchCare merchandise fig P7900B; WatchCare System for VersaCare Bed Rev. A-J merchandise fig P00697901; WatchCare System for VersaCare Bed Rev. K merchandise fig P00697902; WatchCare System for Progressa Bed merchandise fig P00697903; and WatchCare System for Centrella Bed merchandise fig P00697905.
More astir Baxter’s WatchCare system
Baxter acquired the WatchCare strategy arsenic portion of its $12.5 cardinal acquisition of Hillrom successful precocious 2021.
The instrumentality is simply a astute incontinence pad that integrates into healthcare installation beds. It’s meant to supply timely alerts to assistance trim diligent vulnerability to moisture. The Hillrom webpage for the strategy points to probe saying that WatchCare reduced the clip that tegument was exposed to moisture from much than hours to 19 minutes successful a 36-bed med-surg unit.